Revision B
120 Series Maternal/Fetal Monitor
6-9
2015590-001
Functional Checkout Procedure: MECG Test
15. Set the simulator’s
ECG Rate
switch to the MANUAL position and the
Manual
Adjustment
knob to the fully counterclockwise position. Disconnect the ECG
simulator sub-cable from the monitor’s y-adapter cable. Verify the following
on the monitor:
The MHR/P value and mode are both blank.
The recorder stops printing maternal heart rate data on the strip chart paper.
The recorder prints the message
CARDIO INOP
on the center margin of the
strip chart paper after approximately 30 seconds.
16. Set the simulator’s
ECG Mode
switch to the OFF position.
Figure 6-2. MECG Ramp
Содержание Corometrics 126
Страница 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 2: ......
Страница 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Страница 30: ...1 10 120 Series Maternal Fetal Monitor Revision B 2015590 001 Safety Equipment Symbols For your notes ...
Страница 404: ...A 6 120 Series Maternal Fetal Monitor Revision B 2015590 001 Factory Defaults Table of Defaults For your notes ...
Страница 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Страница 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Страница 412: ......
Страница 414: ......
Страница 416: ......
Страница 418: ......
Страница 420: ......
Страница 422: ......
Страница 424: ......
Страница 426: ......
Страница 428: ......
Страница 430: ......
Страница 432: ......
Страница 434: ......
Страница 436: ......
Страница 438: ......
Страница 439: ......