Troubleshooting: MECG Troubleshooting
MECG Troubleshooting
Table 15-7. MECG Troubleshooting
PROBLEM
PROBABLE CAUSE
POSSIBLE SOLUTION
MECG erratic or not functioning properly.
Transducer not properly connected to
monitor.
Electrodes not properly placed.
Clips not attached to electrodes properly.
Electrode gel dried.
Defective MECG cable.
Service required.
Selected lead providing
inadequate signal.
Ensure transducer is securely attached
to monitor.
Re-apply electrodes.
Check clip attachments.
Check electrodes and re-apply gel if
necessary.
Replace cable.
Call Biomedical Engineering
Department.
Change lead selection on MHR/P Setup
screen.
Dashes (– – –) shown in MHR/P area of
display.
Monitor unable to make a
determination due to insufficient signal.
Ensure patient is not asystolic.
Ensure electrodes are firmly secured to
patient.
Содержание Corometrics 126
Страница 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 2: ......
Страница 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Страница 30: ...1 10 120 Series Maternal Fetal Monitor Revision B 2015590 001 Safety Equipment Symbols For your notes ...
Страница 404: ...A 6 120 Series Maternal Fetal Monitor Revision B 2015590 001 Factory Defaults Table of Defaults For your notes ...
Страница 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Страница 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Страница 412: ......
Страница 414: ......
Страница 416: ......
Страница 418: ......
Страница 420: ......
Страница 422: ......
Страница 424: ......
Страница 426: ......
Страница 428: ......
Страница 430: ......
Страница 432: ......
Страница 434: ......
Страница 436: ......
Страница 438: ......
Страница 439: ......