General Maintenance: Maternal Blood Pressure Tests
Patient-to-Line Leakage for IUP
1.
Connect an IUP test body to the monitor’s front panel
UA
input. The IUP test
body can be a SensorTip transducer (wrapped in aluminum foil) or the assembly
leakage cable.
2.
Set the fixture’s
Leakage
switch to
Line
.
3.
Turn ON the monitor using the monitor’s power switch.
4.
Verify the following:
F
< 50
µ
A (100, 120, 220, 230, or 240 VAC monitors)
5.
Record the patient-to-ground leakage for the following conditions:
Table 9-6. Patient-to-Line Leakage Conditions for IUP
Neutral
Ground
Power (Polarity)
Closed
Closed
Normal
Closed
Closed
Reversed
Open
Closed
Normal
Open
Closed
Reversed
Содержание Corometrics 126
Страница 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 2: ......
Страница 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Страница 30: ...1 10 120 Series Maternal Fetal Monitor Revision B 2015590 001 Safety Equipment Symbols For your notes ...
Страница 404: ...A 6 120 Series Maternal Fetal Monitor Revision B 2015590 001 Factory Defaults Table of Defaults For your notes ...
Страница 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Страница 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Страница 412: ......
Страница 414: ......
Страница 416: ......
Страница 418: ......
Страница 420: ......
Страница 422: ......
Страница 424: ......
Страница 426: ......
Страница 428: ......
Страница 430: ......
Страница 432: ......
Страница 434: ......
Страница 436: ......
Страница 438: ......
Страница 439: ......