Revision B
120 Series Maternal/Fetal Monitor
4-17
2015590-001
Theory of Operation: Main Motherboard
61
SP3B/
Output
Spare Chip Select Line
62
+5V
Output
+5 Volt Logic Supply
63
LDSB/
Output
Processor Lower Data Strobe
64
R/WB/
Output
Processor Read/Write Line
65
AGND
Output
Analog Ground
66
DTACK/
I/O
Processor DTACK Transfer Line
67
No Connection
—
—
68
No Connection
—
—
69
CKC
Output
20 MHz Clock
70
GND
Output
Digital Ground
71
No Connection
—
—
72
No Connection
—
—
73
No Connection
—
—
74
No Connection
—
—
75
GND
Output
Digital Ground
76
–15V
Output
–15 Volt Analog Supply
77
AGND
Output
Analog Supply
78
TECG
Input
Telemetry ECG
79
TOCO
Input
TOCO-Telemetry
80
RESB/
Output
Processor Controlled Reset Line RESB/
Table 4-7. Communications Board Connector J5 (Continued)
Pin Number
Signal Name
Signal Type
Signal Description
Содержание Corometrics 126
Страница 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 2: ......
Страница 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Страница 30: ...1 10 120 Series Maternal Fetal Monitor Revision B 2015590 001 Safety Equipment Symbols For your notes ...
Страница 404: ...A 6 120 Series Maternal Fetal Monitor Revision B 2015590 001 Factory Defaults Table of Defaults For your notes ...
Страница 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Страница 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Страница 412: ......
Страница 414: ......
Страница 416: ......
Страница 418: ......
Страница 420: ......
Страница 422: ......
Страница 424: ......
Страница 426: ......
Страница 428: ......
Страница 430: ......
Страница 432: ......
Страница 434: ......
Страница 436: ......
Страница 438: ......
Страница 439: ......