9-12
120 Series Maternal/Fetal Monitor
Revision B
2015590-001
General Maintenance: Maternal Blood Pressure Tests
Electrical Safety Tests
Perform all electrical safety tests in this section.
IMPORTANT
UNITS OF MEASURE—The specified leakage tester (E278-06)
measures in mV. There is a direct correlation of mV to
µ
A. In
other words, on this text fixture, 1 mV = 1
µ
A.
IMPORTANT
220/230/240 VAC POWER CORD—For 220/230/240 VAC
testing, use only P/N 600034 line cord to power the E278-06 test
fixture.
Initial Conditions
1.
Attach the monitor power cord to the leakage test fixture.
2.
Verify that the leakage test fixture’s main AC power cord is attached.
AC Line
F
Voltage measurement from 100–240 V.
Ground Impedance
F
< 300 m
Ω
, lead to rear panel equipotential ground lug
F
< 300 m
Ω
, lead to top cover bezel screw
Содержание Corometrics 126
Страница 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 2: ......
Страница 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Страница 30: ...1 10 120 Series Maternal Fetal Monitor Revision B 2015590 001 Safety Equipment Symbols For your notes ...
Страница 404: ...A 6 120 Series Maternal Fetal Monitor Revision B 2015590 001 Factory Defaults Table of Defaults For your notes ...
Страница 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Страница 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Страница 412: ......
Страница 414: ......
Страница 416: ......
Страница 418: ......
Страница 420: ......
Страница 422: ......
Страница 424: ......
Страница 426: ......
Страница 428: ......
Страница 430: ......
Страница 432: ......
Страница 434: ......
Страница 436: ......
Страница 438: ......
Страница 439: ......