12-28
120 Series Maternal/Fetal Monitor
Revision B
2015590-001
Upgrading a 120 Series Monitor: MECG Option Upgrade Kit
MECG Option Upgrade Kit
The MECG Option Upgrade Kit, adds MECG monitoring to Model 126/126F/128/
128F Monitors. Installation instructions are provided in the kit; they are repeated
here for informational purposes.
Unpacking
Unpack the upgrade kit and ensure it contains all items on the packing list. If
something is missing, contact your GE Service Representative immediately.
Installation
1.
Turn off the 120 Series Monitor and disconnect the power cord from the
monitor.
2.
Remove the nine screws which secure the monitor top cover. Four screws are
located on the bottom of the monitor and five screws are located in the back.
3.
Remove the cover by sliding it toward the rear of the monitor.
4.
Remove the hold-down cover for the front end boards. Three screws secure the
top; one screw is located on the side panel.
5.
Remove the UA/FECG Board from the Front End Motherboard. The UA/FECG
Board is the second board from the
right
in the front end section. (Refer to
Figure 12-22
.)
6.
Remove the MECG Board from its anti-static container.
7.
Mount the MECG Board onto the UA/FECG Board at the headers labeled J5
and J6. (Refer to
Figure 12-1
.)
8.
Secure the MECG Board with the two screws provided in the kit. One screw is
placed at the top left; the other at the bottom right. (See
Figure 12-1
.)
9.
Re-insert the UA/FECG Board (with piggybacked MECG Board) into slots J5
and J8 on the Front End Motherboard. (Refer to
Figure 12-24
.)
10. Connect the front panel MECG harness cable to J7 on the MECG Board. (Refer
to
Figure 12-3
.)
11. The MECG harness cable is already connected to the front panel in all 120
Series Monitors.
12. Replace the hold-down cover for the front end boards and secure with all four
screws.
13. Locate switch pack SW1 on the Main Motherboard and ensure switch 8 is set to
the OFF position. (See
Figure 12-26
.) This switch activates communication
between the Main Motherboard and the MECG Board.
14. Replace the top cover and secure all nine screws.
15. Disconnect all cables from the monitor’s front panel connectors.
16. Re-connect the monitor power cord and turn on the monitor.
Содержание Corometrics 126
Страница 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 2: ......
Страница 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Страница 30: ...1 10 120 Series Maternal Fetal Monitor Revision B 2015590 001 Safety Equipment Symbols For your notes ...
Страница 404: ...A 6 120 Series Maternal Fetal Monitor Revision B 2015590 001 Factory Defaults Table of Defaults For your notes ...
Страница 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Страница 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Страница 412: ......
Страница 414: ......
Страница 416: ......
Страница 418: ......
Страница 420: ......
Страница 422: ......
Страница 424: ......
Страница 426: ......
Страница 428: ......
Страница 430: ......
Страница 432: ......
Страница 434: ......
Страница 436: ......
Страница 438: ......
Страница 439: ......