Revision B
120 Series Maternal/Fetal Monitor
15-7
20155
90
-001
Troubleshooting: Fetal Pulse Oximetry Troubleshooting
Sensor Lifted icon
displayed in
FSpO
2
area of display.
Sensor not making adequate contact at
the sensor site on the fetus.
Damaged sensor.
Assess sensor position relative to fetal
head. Assess sensor depth relative to
the fetal presenting part; advance or
withdraw the sensor as indicated in
“FSpO2 Status Icons”
on page 5-20.
If the icon does not respond to
advancing or withdrawing the sensor,
remove the sensor and place a new
sensor, as directed in
“FSpO2 Status
Icons”
on page 5-20.
Sensor Disconnected icon
displayed in FSpO
2
area of display.
Sensor disconnected from fetal patient
module cable; or fetal patient module
cable disconnected from monitor.
Damaged fetal patient module.
Invalid FSpO
2
sensor connected to fetal
patient module cable.
Ensure fetal patient module cable is
firmly attached to sensor assembly and
monitor. If connections are OK, use a
new sensor.
Replace with another Corometrics Fetal
Patient Module.
Replace with valid Nellcor OxiFirst Fetal
Oxygen Sensor (Series FS14).
COMM
message shown in FSpO
2
area of
display.
Communication error between the built-
in FSpO
2
module and the remainder of
the monitor circuitry.
Contact Biomedical Engineering
Department.
FHR slows during, or immediately after,
sensor placement.
Changes in intracranial pressure due to
examination could cause a reflex
bradycardia stimulus.
Stop the exam. Do not proceed with
sensor placement. Wait for the FHR to
return to the previous range.
Dashes (– – –) shown in FSpO
2
display
area.
Monitor unable to make a determination
due to insufficient signal.
Improperly applied sensor.
Excessive fetal movement.
Excessive maternal movement.
Maternal uterine contraction.
Damaged sensor.
SLOW mode selected.
Check centimeter mark on stylet
chamber relative to the introitus. The
sensor may have ascended or
descended independent of fetal head
descent.
Ensure that the fetal patient module
cable is firmly attached to the monitor
and to the sensor assembly.
Move sensor to another location.
Wait for movement to stop.
Restrict patient movement.
Wait for contraction to end.
Replace sensor.
Access FSpO
2
Setup screen and select
FAST mode.
REPAIR
message shown in FSpO
2
display
area.
System error or self-test failure.
Contact Biomedical Engineering
Department.
Table 15-4. Fetal Pulse Oximetry Troubleshooting (Continued)
Problem
Probable Cause
Possible Solution
)
Содержание Corometrics 126
Страница 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 2: ......
Страница 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Страница 30: ...1 10 120 Series Maternal Fetal Monitor Revision B 2015590 001 Safety Equipment Symbols For your notes ...
Страница 404: ...A 6 120 Series Maternal Fetal Monitor Revision B 2015590 001 Factory Defaults Table of Defaults For your notes ...
Страница 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Страница 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Страница 412: ......
Страница 414: ......
Страница 416: ......
Страница 418: ......
Страница 420: ......
Страница 422: ......
Страница 424: ......
Страница 426: ......
Страница 428: ......
Страница 430: ......
Страница 432: ......
Страница 434: ......
Страница 436: ......
Страница 438: ......
Страница 439: ......