12-34
120 Series Maternal/Fetal Monitor
Revision B
2015590-001
Upgrading a 120 Series Monitor: Software Version 3.5
Software Version 3.5
You can upgrade to software version 3.5 by using the Flasher Software Utility or by
replacing the EEPROMS. Refer to
“Chapter 14, Replacement Parts”
for kit
numbers.
Flasher Software Utility Upgrade Method
Flasher is a software utility program which uses one of the monitor’s RS-232 serial
ports to upgrade to a newer software release. Each Flasher disk contains the software
upgrade for one-time use only. (In other words, you need an individual Flasher disk
for each monitor being upgraded.) The kit comes with an instruction manual for
running the Flasher program.
EEPROM Replacement Method
1.
Turn off the 120 Series Monitor and disconnect the power cord from the
monitor.
2.
Remove the nine screws which secure the monitor top cover. Four screws are
located on the bottom of the monitor and five screws are located in the back.
3.
Disconnect the display interconnect cable from the DSP Board at J5. (Refer to
Figure 12-30
and
Figure 12-1
.)
4.
Disconnect the membrane switch panel ribbon cable from the DSP Board at J4.
(Refer to
Figure 12-30
and
Figure 12-1
.)
5.
Disconnect the blood pressure tubing from the nipples located at the divider to
the recorder cavity. Unthread the tubing from the notch at the top of the DSP
Board. Leave the other end of the tubing attached to the Blood Pressure Board.
(Refer to
Figure 12-30
.)
6.
Remove the two screws which fasten the Blood Pressure Board to the monitor
chassis.
7.
Remove the Blood Pressure Board and set aside on an anti-static surface.
8.
Remove the two screws which fasten the DSP Board to the monitor chassis.
9.
Remove the DSP Board and set aside on an anti-static surface.
10. Locate and note the orientation of flash ROMs U31 and U32 on the Main
Board. (Refer to
Figure 12-31
.) Using the PLCC extractor, carefully remove
U31 and U32.
11. Locate and note the orientation of flash ROMs U8 and U9 on the DSP Board.
(Refer to
Figure 12-1
.) Using the PLCC extractor, carefully remove U8 and U9.
NOTE:
Each flash ROM in the kit is labeled with an assembly number, software
release date, and lot code. Use the assembly number for identification
purposes during this procedure.
12. Remove the four flash ROMs from their anti-static containers and mount them
according to
Table 12-1
.
Содержание Corometrics 126
Страница 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 2: ......
Страница 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Страница 30: ...1 10 120 Series Maternal Fetal Monitor Revision B 2015590 001 Safety Equipment Symbols For your notes ...
Страница 404: ...A 6 120 Series Maternal Fetal Monitor Revision B 2015590 001 Factory Defaults Table of Defaults For your notes ...
Страница 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Страница 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Страница 412: ......
Страница 414: ......
Страница 416: ......
Страница 418: ......
Страница 420: ......
Страница 422: ......
Страница 424: ......
Страница 426: ......
Страница 428: ......
Страница 430: ......
Страница 432: ......
Страница 434: ......
Страница 436: ......
Страница 438: ......
Страница 439: ......