Revision B
120 Series Maternal/Fetal Monitor
3-19
2015590-001
Controls, Indicators, and Connectors: Rear Panel Description
CAUTION
NON-DESTRUCTIVE VOLTAGE—The maximum non-
destructive voltage that may be applied to the rear panel
connectors is 0 volts. Do not attempt to connect cables to these
connectors without contacting your Biomedical Engineering
Department or
Information Technologies
Service Representative.
This is to ensure the connectors comply with leakage-current
requirements of one of the following applicable standards:
Underwriters Laboratories UL-2601.1, Canadian Standards
Associations CSA 22.2 No. 125, or International Electrotechnical
Commission EN60601.1.
L
Equipotential Lug
A binding post terminal is directly connected to the
chassis for use as an equipotentiality connection.
M
AC Voltage Selection
Switch
This switch is intended for qualified service
personnel to select a voltage range for the AC
input:
120: Accepts an AC input in the range of 100–
120 VAC.
240: Accepts an AC input in the range of 220–
240 VAC
N
Power Entry Module
AC line power cord receptacle. Refer to the rear
panel markings to verify line voltage and line
frequency requirements.
Table 3-13. 120 Series Rear Panel (Standard and Optional Features)
Name
Description
NOTE:
Although the J104 Nurse
Call connector is physically present
on the optional communications
package, this connector is only
supported as part of the Spectra
Alerts option.
Содержание Corometrics 126
Страница 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 2: ......
Страница 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Страница 30: ...1 10 120 Series Maternal Fetal Monitor Revision B 2015590 001 Safety Equipment Symbols For your notes ...
Страница 404: ...A 6 120 Series Maternal Fetal Monitor Revision B 2015590 001 Factory Defaults Table of Defaults For your notes ...
Страница 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Страница 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Страница 412: ......
Страница 414: ......
Страница 416: ......
Страница 418: ......
Страница 420: ......
Страница 422: ......
Страница 424: ......
Страница 426: ......
Страница 428: ......
Страница 430: ......
Страница 432: ......
Страница 434: ......
Страница 436: ......
Страница 438: ......
Страница 439: ......