7-18
120 Series Maternal/Fetal Monitor
Revision B
2015590-001
Calibration: Display Check
13. If you suspect that either the Display Driver Board or the Controller Power Unit
are not functioning correctly and you have another properly functioning 120
Series Monitor, substitute the entire display assembly and check for proper
operation. Refer to
“Removing the Display Assembly”
and
“Replacing the
Display Assembly”
. Contact your GE Service Representative for ordering
information.
Removing the Display Assembly
1.
Turn off the 120 Series Monitor and disconnect the power cord from the
monitor.
2.
Remove the nine screws which secure the monitor top cover. Four screws are
located on the bottom of the monitor and five screws are located in the back.
3.
Remove the cover by sliding it toward the rear of the monitor.
4.
Disconnect the display interconnect cable from the DSP Board at J5.
5.
Disconnect the ribbon cable from the Recorder Board at J2. Leave the other
end connected to the Main Motherboard.
6.
Disconnect the Power Supply cable harness from the Recorder Board at J1.
7.
Disconnect the ribbon cable from the membrane switch panel. Leave the other
end connected to the DSP Board.
8.
Remove the four screws which fasten the front bezel to the monitor—two on
each side.
9.
Gently tilt the bezel forward until it lays flat. Be careful not to disconnect the
front end harness cable ferrite beads from their clips. The number of beads
varies depending on which options are installed in your monitor.
10. Remove the four screws which attach the display shield to the front bezel.
Retain the screws.
11. Set aside, but retain, the display shield.
12. Disconnect the display interconnect cable from J1 on the display Controller
Power Unit. Retain this cable.
13. Remove the four standoffs at the corner of the display. Retain the standoffs.
Leave the washers and spacers in place on the studs.
Содержание Corometrics 126
Страница 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 2: ......
Страница 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Страница 30: ...1 10 120 Series Maternal Fetal Monitor Revision B 2015590 001 Safety Equipment Symbols For your notes ...
Страница 404: ...A 6 120 Series Maternal Fetal Monitor Revision B 2015590 001 Factory Defaults Table of Defaults For your notes ...
Страница 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Страница 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Страница 412: ......
Страница 414: ......
Страница 416: ......
Страница 418: ......
Страница 420: ......
Страница 422: ......
Страница 424: ......
Страница 426: ......
Страница 428: ......
Страница 430: ......
Страница 432: ......
Страница 434: ......
Страница 436: ......
Страница 438: ......
Страница 439: ......