13-4
120 Series Maternal/Fetal Monitor
Revision B
2015590-001
Technical Specifications: Operating Modes
MECG Mode
Technique:
Maternal ECG Electrode Type:
Leads Available:
Heart Rate Counting Range:
Heart Rate Resolution:
Heart Rate Averaging:
Heart Rate Update Rate:
Countable Input Signal Range:
Baseline Drift:
Tall T-wave Rejection:
Heart Rate Meter Response Time:
80–120 BPM Step Increase:
80–40 BPM Step Decrease:
Alarm Time for Tachycardia 80–200 BPM:
Offset Voltage Tolerance (Differential):
Maximum Common Mode Voltage:
Preamplifier Bandwidth:
Common Mode Rejection:
Balanced:
Unbalanced 5K RA or LA:
Input Equivalent Noise:
Input Impedance:
Differential:
Common Mode:
Mains Frequency Rejection:
Leakage Current:
Isolation, Mains-to-Patient:
Leads Off Detection:
Alarms:
Audio:
Visual:
Limits:
Technical:
Tachycardia Response Time:
Pacemaker Detection/Rejection:
Input Voltage Range:
Input Pulse Width:
Pulse Rise/Fall Time:
Overshoot/Undershoot:
Peak detecting, beat-to-beat cardiotachometer
Medtronic 1700-003 or equivalent
I, II, and III
30–240 BPM
±1 BPM
1 second average
> 1 update per second
< 0.5 mV to 5 mV peak-to-peak
< 0.5 mV RTI
< 0.8
x
QRS amplitude
< 2 seconds
< 3 seconds
< 10 seconds (high alarm limit at 100 BPM)
±300 mVdc maximum
20 V peak-to-peak
0.6 to 40 Hz
> 80 dB at mains frequency, with patient cable
> 50 dB at mains frequency
< 30 µV peak-to-peak
> 2.5 M
Ω
> 10 M
Ω
> 40 dB
< 60 µA at 254 VAC, with cable, electrically isolated
> 4 kVAC
dc current < 0.1 µA
Alternating 1.5 second chimes (773 Hz and 523 Hz)
Flashing heart rate numeric or message
User-selectable high and low maternal heart rate
Leads off
< 8 seconds
±2.5 mV to ±700mV
0.1 to 2 ms
< 10% of pulse width; not greater than 100 µs
2 mV (see caution below)
Table 13-2. Operating Mode Specifications (Continued)
CAUTION
Excessive overshoot time of pacemaker pulse may cause false
QRS detection.
Содержание Corometrics 126
Страница 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 2: ......
Страница 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Страница 30: ...1 10 120 Series Maternal Fetal Monitor Revision B 2015590 001 Safety Equipment Symbols For your notes ...
Страница 404: ...A 6 120 Series Maternal Fetal Monitor Revision B 2015590 001 Factory Defaults Table of Defaults For your notes ...
Страница 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Страница 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Страница 412: ......
Страница 414: ......
Страница 416: ......
Страница 418: ......
Страница 420: ......
Страница 422: ......
Страница 424: ......
Страница 426: ......
Страница 428: ......
Страница 430: ......
Страница 432: ......
Страница 434: ......
Страница 436: ......
Страница 438: ......
Страница 439: ......