12-18
120 Series Maternal/Fetal Monitor
Revision B
2015590-001
Upgrading a 120 Series Monitor: Nellcor MSpO
2
and NBP Option Upgrade Kit
16. Guide the metal tabs of the new membrane switch panel through the front panel.
Secure the switch panel to the bezel with the nuts retained in
step 12
. Tighten to
8 in-lb.
17. Ensure that the Recorder LED Board is connected to the membrane switch
panel. Using the screws and washers retained in 13, fasten the switch panel
metal grounding tabs to the Recorder LED Board. Tighten to 4 in-lb.
18. Re-connect the Trim Knob cable assembly to the six pin connector on the
membrane switch panel.
19. Ensure the front end harness cable ferrite beads are inserted in the
corresponding clips. (Refer to
Figure 12-6
.)
20. Replace the front bezel and secure with all four screws.
21. Re-connect the ribbon cable from the DSP Board to the membrane switch
panel.
22. Re-connect the Power Supply harness cable to J1 on the Recorder Board.
23. Re-connect the ribbon cable from the Main Motherboard to J2 on the Recorder
Board.
24. Re-connect the display interconnect cable to the DSP Board at J5.
25. Remove the adhesive backing from the overlay label and carefully align over
the front panel membrane switches. (Refer to
Figure 12-7
.)
26. Remove the Blood Pressure Board from its anti-static container.
27. Insert the Blood Pressure Board into the slot labeled J4 on the Main
Motherboard. (Refer to
Figure 12-8
.)
28. Secure the board to the chassis using the two #6-32 x 3/8 screws provided in the
kit.
29. Route the blood pressure tubing through the notch at the top of the DSP Board.
(See
Figure 12-8
.)
30. Attach the tubes to the nipples located at the divider to the recorder cavity. (See
Figure 12-8
.)
31. Remove the hold-down cover for the front end boards. Three screws secure the
top; one screw is located on the side panel.
32.
If FSpO
2
is already installed
in your monitor, disconnect the FSpO
2
harness
cable from the FSpO
2
Board at J1. Remove the Oximeter Carrier Board (No.
13551) from the Front End Motherboard; this board occupies the second slot
from the left while facing the front of the monitor.
If FSpO
2
is not installed
in your monitor, remove the MSpO
2
Board and the
Oximeter Carrier Board from their anti-static containers. If FSpO
2
is installed in
your monitor,
you already have an Oximeter Carrier Board.
NOTE:
The Oximeter Carrier Board is designed so that the MSpO
2
Board mounts
on one side, while the FSpO
2
Board (if installed) mounts on the other side.
The MSpO
2
Board mounts on the
component
side—with the two ribbon
cables.
Содержание Corometrics 126
Страница 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 2: ......
Страница 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Страница 30: ...1 10 120 Series Maternal Fetal Monitor Revision B 2015590 001 Safety Equipment Symbols For your notes ...
Страница 404: ...A 6 120 Series Maternal Fetal Monitor Revision B 2015590 001 Factory Defaults Table of Defaults For your notes ...
Страница 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Страница 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Страница 412: ......
Страница 414: ......
Страница 416: ......
Страница 418: ......
Страница 420: ......
Страница 422: ......
Страница 424: ......
Страница 426: ......
Страница 428: ......
Страница 430: ......
Страница 432: ......
Страница 434: ......
Страница 436: ......
Страница 438: ......
Страница 439: ......