7-24
120 Series Maternal/Fetal Monitor
Revision B
2015590-001
Calibration: Nautilus Tocotransducer Calibration
Equipment Required
You will need the following:
Phillips Head Torque Driver
Screw Cap Extraction Tool
Digital Voltmeter: DM501 or equivalent with x1 probe or leads
Common Mode Rejection (CMR) Test Jack (
Figure 7-4
)
Potentiometer Adjustment Tool
52.5 gram weight
NOTE:
You will need a specific weight to calibrate the Nautilus
Tocotransducer. Refer to
“Chapter 14, Nautilus Tocotransducer”
for
part number information.
Torque Seal
Loctite Adhesive #454 (recommended)
Anti-Static Wristband
Static-Free Work Surface
Calibrated 120 Series Monitor
Figure 7-6. Sealing Screw Caps
Figure 7-7. Nautilus Tocotransducers—Cover Removed
Flat Screw Cap
Button-Style
Loop-Style
2264 GAX/HAX/JAX/KAX/LAX/MAX
R11
Gain
R6
Offset
R15
Gain
R5
Offset
R6
Offset Trim
2264 AAX/BAX/CAX/DAX/EAX/FAX
Содержание Corometrics 126
Страница 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 2: ......
Страница 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Страница 30: ...1 10 120 Series Maternal Fetal Monitor Revision B 2015590 001 Safety Equipment Symbols For your notes ...
Страница 404: ...A 6 120 Series Maternal Fetal Monitor Revision B 2015590 001 Factory Defaults Table of Defaults For your notes ...
Страница 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Страница 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Страница 412: ......
Страница 414: ......
Страница 416: ......
Страница 418: ......
Страница 420: ......
Страница 422: ......
Страница 424: ......
Страница 426: ......
Страница 428: ......
Страница 430: ......
Страница 432: ......
Страница 434: ......
Страница 436: ......
Страница 438: ......
Страница 439: ......