Self-Tests: Diagnostic Control Screen
NO ADDRESS
Wallplate cable may be disconnected; or Corolan address checking is
disabled.
NO DATA YET
Contact GE Service Representative.
NEVER INTERRUPTED
There is no communication with the Corolan master device (e.g. QS Server).
The master device is not polling the monitor.
NO RECENT INTERRUPT
The monitor was successfully communicating with the Corolan master
device (e.g. QS Server); however, two minutes have passed since activity.
DRIVER NOT RUN
Contact GE Service Representative.
TRANSMITTING
LONG TRANSMISSION
IN TEST MODE
NO RECENT RECEPTION
PACKET ERROR
PACKET TOO SHORT
SIM RECEIVED
DISC RECEIVED
OFFLINE ADDRESS CONTENTION
Two devices (monitors) have the same wallplate address. Both devices
have automatically been taken offline.
WAS OFFLINE
Contact GE Service Representative.
HAD DISC ADDRESS
CRC ERROR
UNKNOWN PACKET TYPE
RECENT ADDRESS CONTENTION
A monitor has just been turned on using the same wallplate address as
another monitor. If the situation continues, this message will change to
OFFLINE ADDRESS CONTENTION. Disconnect both monitors from
wallplate. Check wallplate addresses. Re-connect.
SOME ADDRESS CONTENTION
PACKET NOT A POLL
Contact Corometrics Service Representative.
NO NETWORK ACTIVITY
Twenty seconds have elapsed since last network activity detected.
15 SEC SINCE ACTIVITY
Fifteen seconds have elapsed since last network activity detected.
10 SEC SINCE ACTIVITY
Ten seconds have elapsed since last network activity detected.
5 SEC SINCE ACTIVITY
Five seconds have elapsed since last network activity detected.
AWAITING MANAGER DISCONNECT
Contact Corometrics Service Representative.
NO SNRM YET
Table 8-3. Corolan Network Status Messages
Message
Meaning
Содержание Corometrics 126
Страница 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 2: ......
Страница 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Страница 30: ...1 10 120 Series Maternal Fetal Monitor Revision B 2015590 001 Safety Equipment Symbols For your notes ...
Страница 404: ...A 6 120 Series Maternal Fetal Monitor Revision B 2015590 001 Factory Defaults Table of Defaults For your notes ...
Страница 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Страница 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Страница 412: ......
Страница 414: ......
Страница 416: ......
Страница 418: ......
Страница 420: ......
Страница 422: ......
Страница 424: ......
Страница 426: ......
Страница 428: ......
Страница 430: ......
Страница 432: ......
Страница 434: ......
Страница 436: ......
Страница 438: ......
Страница 439: ......