4-40
120 Series Maternal/Fetal Monitor
Revision B
2015590-001
Theory of Operation: Communications Board
Model 2116B Keyboard Interface
The keyboard interface consists of a data latch (U7), a status receive buffer (U8), a
control latch (U6), and a PAL device (U15). Data is latched into U7 by an inverted
version of the incoming STB* signal. This signal is inverted in the PAL device.
The STB* signal is buffered through U8 before going to the PAL device for static
protection. Once the STB* signal goes low, the PAL device IBF output goes high
(tells keyboard that the monitor cannot accept any more data). After the STB* line
goes high, the PAL-generated DATAVA* signal (goes to U8 status port) is read by
the processor signaling that data is available. The processor then reads the data by
pulsing the DATARD* line to U7. The PAL device then sets the DATAVA* line
high. The IBF line to the keyboard goes low (signaling to the keyboard that the
monitor can accept more data) when the processor clears the IBF latch. This gives
the processor time to read the data and activate the INVLD* line if the nibble was
bad before the next nibble is sent. Control information is sent to the keyboard via
three lines from latch U6. A KBDRST* line resets the keyboard, a KBDBUSY*
tells the keyboard to put up a busy message (due to recorder being off or busy
printing previous message), and a KBDINVLD* tells the keyboard that the monitor
received an invalid message. The status lines from the keyboard consists of a
KBDIN* line which signals that the keyboard is plugged in, and a KBDMARK* line
which pulses for a one-second minimum to print an event mark. These lines are read
through buffer U8.
The +5 V power for the keyboard is created from VR1 using the +12 V source. This
is done for static protection and for reliability reasons.
Model 340 Telemetry Interface
The telemetry interface consists of eight digital input status/mode lines and four
analog inputs. The analog lines (TECG, TMECG, TELMAUDIO, TTOCO) are all
passed to the Main Motherboard which in turn feed the DSP Board and Ultrasound
Board. The eight status lines are buffered and read by the main processor.
Содержание Corometrics 126
Страница 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 2: ......
Страница 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Страница 30: ...1 10 120 Series Maternal Fetal Monitor Revision B 2015590 001 Safety Equipment Symbols For your notes ...
Страница 404: ...A 6 120 Series Maternal Fetal Monitor Revision B 2015590 001 Factory Defaults Table of Defaults For your notes ...
Страница 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Страница 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Страница 412: ......
Страница 414: ......
Страница 416: ......
Страница 418: ......
Страница 420: ......
Страница 422: ......
Страница 424: ......
Страница 426: ......
Страница 428: ......
Страница 430: ......
Страница 432: ......
Страница 434: ......
Страница 436: ......
Страница 438: ......
Страница 439: ......