Revision B
120 Series Maternal/Fetal Monitor
8-19
2015590-001
Self-Tests: J102 Analog Output Connector DAC Static Test
HR1 Mode
Use the Trim Knob control to select: – – –, FECG, OFF, US, or INOP.
HR2 Mode
Use the Trim Knob control to select: TEST, FECG, INOP, US, or MECG.
UA Mode
Use the Trim Knob control to select: – – –, INOP, OFF, IUP, or TOCO.
Markout*, Check Paper*, FMD1, and FMD2
Use the Trim Knob control to alternate between ON and OFF.
NOTE:
Signal names followed by an asterisk (*) or slash (/) are active low.
Calibration
For calibration, you must use the specific values listed in
Table 8-4
(Corometrics
output levels) or
Table 8-5
(Hewlett-Packard output levels). This tests the high and
low ranges for the HR1, HR2, and UA signals. If a measured value does not fall
within the given range, adjust the corresponding potentiometer accordingly (on
Communications Board, No. 13388 or 15297).
Figure 8-5. J102 Analog Output Connector Screen
Содержание Corometrics 126
Страница 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 2: ......
Страница 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Страница 30: ...1 10 120 Series Maternal Fetal Monitor Revision B 2015590 001 Safety Equipment Symbols For your notes ...
Страница 404: ...A 6 120 Series Maternal Fetal Monitor Revision B 2015590 001 Factory Defaults Table of Defaults For your notes ...
Страница 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Страница 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Страница 412: ......
Страница 414: ......
Страница 416: ......
Страница 418: ......
Страница 420: ......
Страница 422: ......
Страница 424: ......
Страница 426: ......
Страница 428: ......
Страница 430: ......
Страница 432: ......
Страница 434: ......
Страница 436: ......
Страница 438: ......
Страница 439: ......