Functional Checkout Procedure: Ultrasound Test
Ultrasound Test
This portion of the functional checkout procedure ensures the integrity of the
ultrasound circuitry and the heart rate channel of the recorder.
1.
Connect the simulator’s US sub-cable to the
US
receptacle on the monitor.
2.
Set the switches on the Model 325 Input Simulator according to
Table 6-3
.
3.
If not already on, depress the monitor’s
Record
pushbutton.
4.
Turn the simulator’s
Manual Adjustment
knob to input an ultrasound signal of
approximately 120 BPM. Verify the following on the monitor:
The FHR1 value is 120 BPM.
The FHR1 mode is
US
.
The FHR1 heartbeat indicator (
) flashes at a rate of 120 times per minute.
Ultrasound audio volume from the rear panel speaker can be increased or
decreased using the left pair of
Volume
pushbuttons. (Set the volume to
the desired level.)
The recorder prints a continuous line at 120 BPM on the top grid of the
strip chart paper.
The recorder prints the message
US
on the center margin of the strip chart
paper after approximately 20 seconds.
5.
Use the simulator’s
Manual Adjustment knob
to increase the heart rate value
by less than 13 BPM from the 120 BPM baseline. Verify the following on the
monitor:
The FHR1 value immediately reflects this new input rate.
The strip chart recorder immediately reflects this new input rate.
6.
Use the simulator’s
Manual Adjustment
knob to decrease the heart rate value
by more than 13 BPM from the 120 BPM baseline. Verify the following on the
monitor:
The FHR1 value immediately reflects this new input rate.
The strip chart recorder prints at the last input rate for an additional 3
seconds before blanking the heart rate data and printing a continuous line at
the new input rate.
7.
Set the simulator’s
US Rate
switch to the RAMP position. Verify that the FHR1
value counts between approximately 50 and 210 BPM and that the recorder
prints a ramp between the same values. (Refer to
Figure 6-6
.)
Содержание Corometrics 126
Страница 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 2: ......
Страница 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Страница 30: ...1 10 120 Series Maternal Fetal Monitor Revision B 2015590 001 Safety Equipment Symbols For your notes ...
Страница 404: ...A 6 120 Series Maternal Fetal Monitor Revision B 2015590 001 Factory Defaults Table of Defaults For your notes ...
Страница 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Страница 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Страница 412: ......
Страница 414: ......
Страница 416: ......
Страница 418: ......
Страница 420: ......
Страница 422: ......
Страница 424: ......
Страница 426: ......
Страница 428: ......
Страница 430: ......
Страница 432: ......
Страница 434: ......
Страница 436: ......
Страница 438: ......
Страница 439: ......