75
20
Syringe barrel clasp
potentiometer
Check the syringe barrel clasp calibration. Run test 7.
See section 6.2.8, page 47.
Calibrate the syringe barrel clasp using Agilia Partner
maintenance software.
If the error remains, check the cable connection to CPU board
(J1 connector). See figure 2.8, page 16.
If the error remains, replace the syringe barrel clasp.
See section 8.7, page 94.
Calibrate the syringe barrel clasp using Agilia Partner
maintenance software and check the pump functionality.
21
Displacement linear
potentiometer
Check the displacement calibration. Run test 6.
See section 6.2.7, page 45.
Calibrate the displacement using Agilia Partner maintenance
software.
If the error remains, check the cable connection to CPU board
(J2 connector). See figure 2.8, page 16.
If the error remains, replace the displacement sensor.
See section 8.12, page 115.
Calibrate the displacement using Agilia Partner maintenance
software and check the pump functionality.
22
Force sensor
Check the force sensor calibration. Run test 5.
See section 6.2.6, page 43.
Calibrate the force sensor using Agilia Partner maintenance
software.
If the error remains, check the cable connection to CPU board
(J9 connector). See figure 2.8, page 16.
If the error remains, replace the force sensor and ribbon cable kit.
See section 8.11, page 110.
Calibrate the force sensor using Agilia Partner maintenance
software and check the pump functionality.
23
Syringe flanges switch
Check the syringe flanges switch is functioning correctly by
running test 7. See section 6.2.8, page 47.
Check the cable connection to CPU board (J1 connector).
See figure 2.8, page 16.
Replace the syringe barrel clasp flexible circuit.
See section 8.7, page 94.
24
Disengagement switch
Check the functionality of the disengagement switch.
Check the cable connection to CPU board (J9 connector).
See figure 2.8, page 16.
Replace the disengagement flexible circuit.
See section 8.10, page 105.
25
Syringe head detection
optical switch
Check the functionality of the syringe head optical switch by
running test 7. See section 6.2.8, page 47.
Check the cable connection to CPU board (J9 connector).
See figure 2.8, page 16.
Replace the plunger kit.See section 8.14, page 121.
26
Board missing
Check the connection between Display board and CPU board.
See section 2.1.3, page 13.
See section 2.1.4, page 16.
Check the connection between Power supply board and CPU
board.
See section 2.1.4, page 16.
See section 2.1.5, page 20.
Replace the faulty part(s).
27
LCD backlight missing
Check the connection between Display board (J3 connector) and
LCD display. See section 2.1.3, page 13.
Replace the LCD display. See section 8.3, page 83.
If the problem cannot be resolved, replace the Display board.
Error Code
Description
Recommended Actions
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...