Warning
Use of the Agilia pump adjacent to or stacked with other
equipment should be avoided because it could result in improper
operation. If such use is necessary, this equipment and the other
equipment should be observed to verify that they are operating
normally.
Use of accessories, transducers and cables other than those
specified or provided by the manufacturer of the Agilia pump
could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result
in improper operation.
Portable RF communications equipment (including peripherals
such as antenna cables, internal and external antennas) should
be used no closer than 10 cm for cell phones and 30 cm for other
equipments, to any part of the Agilia pump, including cables
specified by the manufacturer. Otherwise, degradation of the
essential performances of Agilia pump could result.
Electrosurgical equipment (including base unit, cables,
electrodes) should be used no closer than 30 cm, to any part of
the Agilia pump, including cables specified by the manufacturer.
Otherwise, degradation of the essential performances of Agilia
pump could result.
The user might be required to take mitigation measures, such as
relocating or re-orienting the equipment.
If the Agilia pump is placed near RF communication equipment such as
cell phones, DECT phones or wireless access points, RFID reader &
tags,... It is essential to observe a minimum distance between the Agilia
pump and this equipment specified above. If the Agilia pump causes
harmful interference or if it is itself disrupted, the user is encouraged to
try to correct the interference by one of the following actions:
- Reorient or relocate the Agilia pump, the patient or disruptive
equipment.
- Change the routing of cables.
- Connect the Agilia pump power plug to a protected / backed-up
/ filtered supply or directly to the UPS circuit (uninterruptible
power supply).
- Increase the separation between the Agilia pump and
disruptive equipment.
- Plug the Agilia pump into an outlet on a different circuit from
the one to which the patient or disruptive equipment is
connected.
- In any case, whatever the context, the user should conduct
interoperability testing in a real situation to find the correct
setup and location.
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...