Checking the AC Power Supply / Battery Operation
6.2.24
Electrical Test
Test Description
Run this test to check the power supply indicator state while switching from AC power supply
to battery.
The power supply indicator works properly.
If this test fails, replace the upper case that supports the keypad.
1.
Disconnect the pump’s AC power cord.
2.
Check that the plug-shaped power supply indicator
is off.
3.
Connect the pump to an AC power supply.
4.
Check that the plug-shaped power supply indicator
lights up green.
Test Description
This test checks the electrical safety of the pump according to the IEC 60601-1 medical
electrical equipment standard.
The pump is compliant with the IEC 60601-1 medical electrical equipment standard.
Contact your biomedical department or your
Fresenius Kabi
sales representative.
Required Tools and Equipment
No specific tools or equipment are required
Run the electrical safety test according to the IEC 60601-1 medical electrical equipment standard.
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...