![Fresenius Kabi Agilia SP Скачать руководство пользователя страница 229](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626229.webp)
81
8.15
View Flow Rate History
Symbol
Procedure
Viewing flow rate history
This function allows the user to check the current infusion’s history
information in order to verify the dose administered.
You can view flow rate history as follows:
1.
Press
.
2.
Press the arrow keys to select
.
3.
Press
enter
.
The following information is displayed:
- An event marker (cursor)
- The event details (time and flow rate)
- The measured flow rate (solid line)
4.
Press the
and
buttons to browse the events.
5.
Press
to view information about the selected event.
Information
The history is not refreshed while the history screen is displayed.
To refresh the history data, exit and select the history again.
Flow rate history is not stored after powering off.