13
1.8
Use Environment
Agilia SP MC is intended for use in healthcare facilities, under the
supervision of trained healthcare personnel.
The pump must be used in the following operational conditions to ensure
proper performance:
Operating temperature range:
5 °C to 40 °C
Operating pressure range:
700 hPa (525 mmHg / 10.15 PSI) to 1060 hPa (795 mmHg /
15.37 PSI)
Operating humidity range:
20 % to 90 % with no condensation
Altitude:
Up to 3000 m above sea-level
Warning
The functionality of the pump can be affected by pressure variations,
mechanical shocks, heat ignition sources, and so on.
Information
The pump can be used in ambulances exclusively with the Agilia
Holder Ambulance accessory. Due to use in road ambulances,
performances of the device can be modified. For more
information, refer to Agilia Holder Ambulance IFU.
For more information on using the device in specific conditions,
contact your
Fresenius Kabi
representative.
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...