Unscrew and remove the 2 screws [14].
18.
Remove the syringe barrel clasp bridle [15].
19.
Unscrew and remove the 2 screws [16].
20.
Unclip and remove clip [18].
21.
Pull syringe barrel clasp kit [17] out of its base location.
22.
Disconnect the syringe detector [19] from the syringe barrel clasp kit [17].
23.
Remove the syringe detector [19].
24.
Remove membrane [20].
The syringe detection system [21] comprises syringe detector [19] and membrane [20].
Re-assembling
1.
Perform the dismantling procedure in reverse order to re-assemble the unit.
Do not forget to clip part [18] into the groove of syinge barrel clasp before attaching bridle
[15]. Pay attention to the direction of the flat cable connections when assembling.
2.
Perform the following calibration using Agilia Partner maintenance software:
Syringe barrel clasp
3.
Perform the regular servicing tests. See Quality Control Procedure, page 63.
14
16
14
15
16
15
18
19
21
20
17
20
19
18
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...