7.11.3
Volume Limit
3.
Press the arrow keys to set the infusion time, and press
OK
.
The infusion rate is automatically readjusted.
4.
Press the arrow keys to configure the end of infusion settings and
press
OK
to confirm.
Stop:
The infusion stops when the VTBI is completed.
Keep Vein Open (KVO):
After the VTBI is completed, the infusion
continues at a preset flow rate to keep the access device open.
Continuous:
After the VTBI is completed, the infusion continues
at the programmed flow rate.
5.
Press
start
to start the infusion.
You can use this infusion mode to set a limit to the volume infused.
When the limit is reached, an alarm is triggered.
1.
Access the
Volume Limit
menu, see section 8.12, page 78.
2.
Press the arrow keys to set the volume limit, and press
OK
.
Volume/Time
Dose/Time
Flow rate
Dose