106
13.3
Operations for Syringes
13.3.1
Removing a Syringe
13.3.2
Changing a Syringe
13.3.3
Syringe Replacement Interval
1.
Press
to stop the infusion.
2.
Disconnect the patient.
3.
Open the syringe barrel clasp.
4.
Press
to silence the audible signal for 2 minutes.
5.
Push the disengagement lever down and remove the syringe from
its cradle.
6.
Disconnect the syringe from its extension set.
7.
Disconnect the extension set from the access device in accordance
with healthcare facility protocol.
1.
Remove the syringe,
2.
Prepare a new syringe and follow the steps described in the
flowchart.
3.
After the new syringe is installed, acknowledge the "same therapy"
screen (optional screen).
Warning
Electronically prime the syringe pump system after replacing a near-
empty syringe with a replacement syringe.
Information
Properly dispose of used syringes.
Replace the syringe according to your healthcare facility’s protocol or
CDC guidelines.
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...