![Fresenius Kabi Agilia SP Скачать руководство пользователя страница 42](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626042.webp)
42
6.2.3
Test 2: Maintenance
6.2.4
Test 3: Maintenance Messages
6.2.5
Test 4: Events
Test Description
Run this test to check that the pump maintenance data is correct.
1.
Access the
Tests
menu. See section 6.2.1, page 40.
2.
Press the arrow keys to select
Test 2: Maintenance
.
3.
Press
enter
.
4.
Press the arrow keys to scroll through the information screens
.
The following data is displayed:
Duration
: Running time since last preventive maintenance in hours (if less
than 72 hours), in days (if less than 120 days) or in months
Date
: Last preventive maintenance date
Total duration
: Total running time in hours (if less than 72 hours), in days
(if less than 120 days) or in months.
Total distance
: Total running distance of the plunger’s driver in m.
5.
Press
exit
to return to the
Tests
menu.
Test Description
Run this test to check the maintenance messages entered. The maintenance message is
displayed for information only. It cannot be modified here.
1.
Access the
Tests
menu. See section 6.2.1, page 40.
2.
Press the arrow keys to select
Test 3: Maintenance messages
.
3.
Press
enter
.
Informational messages stored by After Sales Service are displayed.
4.
Press
exit
to return to the
Tests
menu.
Test Description
Run this test to consult the list of pump events, and the details of each recorded technical event.
The clinical events are not displayed here.
1.
Access the
Tests
menu. See section 6.2.1, page 40.
2.
Press the arrow keys to select
Test 4: Events
.
3.
Press
enter
.
The technical events are listed.
4.
Press the arrow keys to select an event.
5.
Press
enter
.
The selected event is displayed in detail.
6.
Press the arrow keys to scroll through the event detail screen.
7.
Press
exit
to return to the list of events.
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...