Installing a Syringe
Data Set information (optional)
Warning
This must only be done when the patient is not connected.
1.
Open the syringe barrel clasp [A].
2.
Push the disengagement lever [B]
down and move the plunger driver to
the right.
3.
Place the syringe in its cradle, with
the flanges correctly inserted in the
provided slot.
4.
Secure the syringe with the syringe
barrel clasp [A].
5.
Push the disengagement lever [B]
and move the plunger driver gently to
the left until it is in contact with the
plunger head.
6.
Check the general installation.
Screen After Powering on
Description
A
B