End of volume limit !
Low (!)
The volume limit is reached and the end of infusion setting is
set as "KVO" or "continuous".
End of volume/time !
Low (!)
The VTBI is completed and the end of infusion setting is set as
"KVO" or "continuous".
End of dose/time !
Low (!)
Pressure increase !
Low (!)
The pressure is increasing in the infusion line.
Check for occlusions in the infusion line.
Drop in pressure !
Low (!)
The pressure is decreasing in the infusion line.
Check the downstream Luer lock connection and the
integrity of the entire line.
High internal temperature !
Low (!)
Temperature increase.
Check device environment.
Send the pump for maintenance.
AC power failure !
Low (!)
The power supply is inconsistent.
Contact your
Technical Service.
Low (!)
If the pump is not used during an extended period, connect to a
power supply and wait until the battery is charged.
Alarm reporting not available
on the Link !
Low (!)
The pump is mounted on a Link+ Agilia rack that has not been
upgraded.
Contact your qualified technician or your
Fresenius Kabi
Technical Service.
Remove completely syringe !
Low (!)
Preventive auto-test on potential failure of plunger head.
Remove and reinstall syringe.
Upper soft max
Information signal
The upper soft limit is exceeded, according to the drug settings
defined in the drug library.
Lower soft min
Information signal
The lower soft limit is exceeded, according to the drug settings
defined in the drug library.
Reached hard limit
Information signal
The upper or lower hard limit is reached.
Power disconnection
Information signal
The pump is disconnected from the AC power. A single beep is
emitted.
Press
to acknowledge.
Check that the battery life is sufficient for the expected
infusion duration.
If the disconnection was unintentional, check the power
connection.
Keypad lock status
Information signal
The keypad is locked.
Unlock the keypad.
Keypad locked
Information signal
The keypad is locked and the syringe barrel clasp was opened
and closed.
Unlock the keypad.
Unlock keypad to continue
Insufficient Dose
Information signal
In TCI mode, the remaining volume/dose in the syringe is
insufficient to reach the target (for Agilia SP TIVA).
High concentration
Information signal
The upper soft limit for concentration is exceeded, according to
the drug settings defined in the drug library (for Agilia SP
TIVA).
Low concentration
Information signal
The lower soft limit for concentration is exceeded, according to
the drug settings defined in the drug library (for Agilia SP
TIVA).
Reached hard limit
Information signal
The upper of lower hard limit for concentration is reached (for
Agilia SP TIVA).
Displayed Message
Priority
Problem / Resolution
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...