10.
Remove the CPU board.
While doing this, take care not to pull out the flat cable on the inside.
11.
Disconnect the flat cable [J9].
12.
Push the disengagement lever down [9] and move the plunger driver to its maximum.
13.
Remove the 2 plastic caps [10] using a flat screwdriver.
14.
Unscrew and remove the 2 screws.
15.
Unclip and remove the syringe flange cradle [11].
16.
Unscrew and remove the 2 screws [12] that hold the base.
17.
Remove the base [13].
While removing the base [13], pay attention to the position of angle bracket clip kit [14]
between the base and the aluminum profile. When re-assembling the base, make sure to
position this part as it was originally.
18.
Unscrew and remove the 3 screws [15] that hold the right flange assembly to the aluminum
profile.
19.
Take the unit apart.