101
12.4.9
Technical Error Alarms
Message
Priority
Stops
Infusion?
Problem / Resolution
Erxx(yyyy) !!!
High (!!!)
Yes
Technical alarm.
Contact your qualified technician or your
Fresenius Kabi
sales representative.
High internal
temperature !
Low (!)
No
Temperature increase.
Check device environment.
Alarm reporting not
available on the
Link !
Low (!)
No
The pump is mounted on a Link+ Agilia rack
that has not been upgraded.
Contact your qualified technician or your
Fresenius Kabi
sales representative.
In the case of a system malfunction, the alarm sounds and an error
message
Erxx(yyyy) !!!
is displayed.
1.
Record the error message
Erxx(yyyy) !!!
.
2.
Disconnect the pump from the power supply.
3.
Reset the device by pressing the
key for 10-15 seconds, and
release it when the device powers off.
Warning
If the alarms persist when the pump is powered on again, do not use the
device on a patient, and contact qualified biomedical engineering staff in
your healthcare facility, or your
Fresenius Kabi
sales representative.
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...