![Fresenius Kabi Agilia SP Скачать руководство пользователя страница 35](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626035.webp)
35
User 12:
Graph items
View flow rate history
Displays or hides menu item: View
flow rate history.
Enabled / Disabled
View pressure history
Displays or hides menu item: View
pressure history.
Enabled / Disabled
View concentration history
Displays or hides menu item: View
concentration history (For
Agilia SP TIVA only).
Enabled / Disabled
User 13:
Default TCI
Numerical screen
Displays or hides menu item: TCI
display (For Agilia SP TIVA only).
Enabled / Disabled
Graphical screen
Enabled / Disabled
Par 1:
Key press sound
Key press sound
Turns the key press sound on or off. Enabled / Disabled
Par 2:
Alarm Volume
Alarm Volume
Sets the volume for alarms.
7 volume levels
Par 3:
Store infusion
parameters
Drug and parameters:
Store drug name only
Store param. infusion
Saves the drug name only, or all
infusion parameters, or nothing.
Enabled / Disabled
Same infusion screen:
Displayed
If
Displayed
is enabled, enables the
display of the same infusion screen.
Enabled / Disabled
Same infusion screen:Time
Duration of context backup.
1
24 hours
Volume infused: Cleared
If
Cleared
is enabled, the infused
volume is reset to 0 at each infusion
start.
Enabled / Disabled
Volume infused: Stored
If
Stored
is enabled, the infused
volume is saved in memory.
Enabled / Disabled
Par 4:
Maximum rate
Maximum rate for 50 mL
syringe
Sets the maximum infusion rate for
50 / 60 mL syringes.
0.1
1200 mL/h
Maximum rate for 30 mL
syringe
Sets the maximum infusion rate for
30 mL syringes.
0.1
600 mL/h
Maximum rate for 20 mL
syringe
Sets the maximum infusion rate for
20 mL syringes.
0.1
600 mL/h
Maximum rate for 10 mL
syringe
Sets the maximum infusion rate for
10 mL syringes.
0.1
350 mL/h
Maximum rate for 5 mL
syringe
Sets the maximum infusion rate for
5 mL syringes.
0.1
250 mL/h
Par 7:
Infusion start
Infusion start: Mandatory
prime
Displays a Mandatory prime
message, an Advised prime
message, or no message prior to
starting an infusion.
Enabled / Disabled
Infusion start: Advised
prime
Enabled / Disabled
Par 8:
Empty syringe
Empty syringe mode
If
Empty syringe mode
is enabled,
an "Empty syringe mode" banner
starts blinking after the
acknowledgement of a medium-
priority "End of infusion" alarm.
Enabled / Disabled
Par 9:
Bolus rates
50 mL
Sets the bolus rate for 50 mL
syringes.
0.1
1200 mL/h
30 mL
Sets the bolus rate for 30 mL
syringes.
0.1
600 mL/h
20 mL
Sets the bolus rate for 20 mL
syringes.
0.1
600 mL/h
10 mL
Sets the bolus rate for 10 mL
syringes.
0.1
350 mL/h
5 mL
Sets the bolus rate for 5 mL
syringes.
0.1
250 mL/h
Par 15:
Syringe/ward name
displays
Syringe name
Displays the syringe or ward name in
the Infusion monitoring screen.
Enabled / Disabled
Ward name
Function
Setting
Description
Range of Settings
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...