![Fresenius Kabi Agilia SP Скачать руководство пользователя страница 66](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626066.webp)
66
6.3.2
Quality Control Certificate
Pump type:
Product code:
Serial number:
Hospital name:
Ward name:
Running time since last
preventive maintenance date:
Date of last preventive maintenance:
Date of next preventive maintenance:
No.
Procedure
Pass
Fail
1
Test 1: Identification and general appearance check.
Pump housing, AC power cord, product code, serial number.
2
Locking system test. Check the attachment lock knob and rotating pole clamp.
3
LCD display test. Check/adjust the LCD contrast (Pump settings > User 3).
4
Test 9: Indication. Backlight, LEDs, indicators and buzzer test.
5
Test 10: Keypad. Keypad test.
6
Test 24: Watchdog.Watchdog.
Default mode triggered in
seconds (between 1.0 second and 2.25 seconds).
7
Test 7: Syringe Barrel Clasp. Check the syringe barrel clasp sensor.
Tool V1 - Diameter: mm (between 9.0 mm and 12.0 mm (0.35 in and 0.47 in)).
Tool V2 - Diameter: mm (between 14 0 mm and 17.0 mm (0.55 in and 0.70 in)).
Tool V4 - Diameter: mm (between 26.5 mm and 29.5 mm (1.04 in and 1.16 in)).
8
Test 5: Force. Occlusion alarm test
Triggering of an occlusion alarm in less than 60 seconds at PSI
(between 8.0 PSI (0.55 bar) and 10.9 PSI (0.75 bar)).
9
Test 6: Displacement calibration test. Checking the displacement sensor
Right stop - Position: mm (between 4.67 in and 5.41 in (118.5 mm and 137.5 mm)).
65 mm gauge - Position: mm (between 2.50 in and 2.62 in (63.5 mm and 66.5 mm)).
Left stop - Position: mm (between 0.39 in and 0.60 in (10.0 mm and 15.2 mm)).
10
Power indicator test. AC power disconnection.
11
Battery life. Test 13: Battery test and Test 12: Battery life.
Displayed "Alarm":
:
(greater than 01h30 (01:30))
12
Electrical tests
Pass
Fail
Name:
Technical department:
Date:
Signature:
Observations:
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...