![Fresenius Kabi Agilia SP Скачать руководство пользователя страница 210](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626210.webp)
62
7.12
Other Functions
7.12.1
Priming the Syringe and the Extension Set
Note: The prime function can be configured with Agilia Vigilant Drug’Lib
(custom profiles) or in the pump options (Basic profile):
Mandatory
: A message is displayed and the user is required to
prime the line before infusion.
Advised
: A message is displayed to encourage the user to prime
the line before infusion.
Not displayed
: The pump does not remind the user to prime the
line before infusion.
Information
We strongly recommend to always use the prime function, especially for
life-sustaining drugs. Priming the syringe and the extension set
removes all the mechanical gaps and ensures that the programmed flow
rate is reached within the smallest delay once the start button is pressed.
1.
Press
to power on the pump.
2.
Press
.
3.
Make sure the extension set is not connected to the patient, as
indicated on the screen.
4.
Press
OK
to proceed.
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...