78
8
Intervention Procedures
This section details all dismantling and re-assembling procedures.
Warning
ONLY use the recommended accessories and spare parts delivered with the pump. NO PART IS
REPAIRABLE. When replacing components, only use
Fresenius Kabi
spare parts. Refer to the
"Spare Parts Catalog" for ordering information.
Any instruments or devices used for maintenance must be regularly checked or re-calibrated
according to their specifications and local regulations.
Disconnect the battery, and discharge the boost capacitor before opening the pump (see
procedure no.1). A risk of electric shock exists when touching a charged boost capacitor.
For safety reasons, the technician should not intervene when the pump is connected to an AC power
supply:
- Unplug the power supply cord from the AC power supply.
- Power off the pump.
When working with electronic components, we recommend wearing an anti-static bracelet
connected to earth and working on an anti-static mat.
Do not touch the electronic boards with your hands.
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...