46
Figure 6.2:
Installation of the 2.56 in (65 mm) displacement gauge
5.
Press the arrow keys to display the
Conversion/Position
screen.
6.
Push the disengagement lever down, and move the plunger driver gently to the right stop.
7.
Check the value displayed for
Position
. The value must be between 4.67 in and 5.41 in
(118.5 mm and 137.5 mm).
8.
Push the disengagement lever down, and move the plunger driver gently to the left stop.
9.
Check the value displayed for
Position
. The value must be between 0.39 in and 0.60 in
(10 mm and 15.2 mm).
10.
Open the syringe barrel clasp and place the 2.56 in (65 mm) gauge in the syringe cradle, with the
flanges correctly inserted in the provided slot and the 2.56 in (65 mm) part positioned to the right.
11.
Secure the gauge with the syringe barrel clasp.
12.
Push the disengagement lever down, and move the plunger driver gently to the left until it is in contact
with the gauge.
13.
Check the value displayed for
Position
. The value must be between
2.5 in and 2.62 in (63.5 mm and 66.5 mm).
14.
Open the syringe barrel clasp.
15.
Push the disengagement lever down and move the plunger driver to the right.
16.
Remove the 2.56 in (65 mm) displacement gauge.
17.
Press
exit
to return to the
Tests
menu.
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...