In the case of a system malfunction, the alarm sounds and an error message
Erxx(yyyy) !!!
is
displayed.
1.
Record the error message
Erxx(yyyy) !!!
.
2.
Disconnect the pump from the power supply.
3.
Reset the device by pressing the
key for 10-15 seconds, and release it when the device powers
off.
Warning
If the alarms persist when the pump is powered on again, do not use the device on a patient, and contact
qualified biomedical engineering staff in your healthcare facility, or your
Fresenius Kabi
sales
representative.
Alarm Priority
Required Operator Response
Description
High (!!!)
Immediate response
The infusion stops.
An alarm description is displayed on the
pump screen.
Medium (!!)
Prompt response
The infusion continues.
Low (!)
Awareness
Information
Signals
Awareness
The infusion continues.
An information message is displayed on the
pump screen.