Procedure #8: Syringe Detection System
Required Tools and Equipment
1 Torx T-10 screwdriver
1 Torx T-8 screwdriver
1 anti-static bracelet
Maintenance level
Level 3 specialist technician.
Prerequisites
The battery is disconnected and the boost capacitor discharged.
See section 8.1, page 79.
Dismantling
1.
Turn the pump upside down.
2.
Unscrew and remove the 3 screws [1].
3.
Remove the angle bracket [2], being careful not to pull out the CPU board flat cable [5].
4.
Disconnect CPU board flat cable [5] at [J1].
5.
Unscrew the 2 screws [3] that hold the upper case [4] to the base.
6.
Remove the upper case [4], being careful not to pull out the display board flat cable.
1
1
1
2
J1
3
4
3
5
Содержание Agilia SP