59
7.11
Infusion Modes
7.11.1
Simple Rate
7.11.2
Volume/Time & Dose/Time
You can program an infusion with the different infusion modes available,
depending on the pump configuration, and on the selected drug.
1.
Press the arrow keys to select the infusion rate.
2.
Press
OK
.
For more information, see section 7.6, page 41.
You can use this infusion mode to program a volume to be infused
(VTBI) over a programmed period of time.
When the VTBI is reached, an alarm is triggered.
1.
For information on how to access the
Volume/Time
or
Dose/Time
menu, see section 8.11, page 78.
2.
Press the arrow keys to set the volume or dose to be infused (VTBI /
DTBI), and press
OK
.
The infusion rate is automatically calculated.
Information
If you program a volume to be infused that is greater than the actual
volume in the syringe, make sure to replace the syringe when it is empty,
Volume/Time
Dose/Time
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...