32
5
Basic Profile Configuration
5.1
Accessing the Basic Profile Configuration
5.2
Basic Profile Configuration Functions
The Basic Profile Configuration menu is available only if Basic Profile has previously been selected as the
default profile. See section 4, page 30.
1.
Access the
Options
menu. See section 3.3, page 29.
2.
Press the arrow keys to select
Basic Profile configuration
.
3.
Press
enter
.
4.
Enter the Basic Profile configuration access code "0200".
5.
Press
OK
.
6.
Press the arrow keys to select the setting category (
User
or
Ward
).
7.
Press
enter
.
Function
Setting
Description
Range of Settings
User 1:
Screen options
Vol-dose information
Displays the volume-dose on the
infusion monitoring screen.
Enabled / Disabled
Time information
Displays the remaining infusion time
on the infusion monitoring screen.
Enabled / Disabled
Battery life information
Displays the the battery life
information on the infusion
monitoring screen.
Enabled / Disabled
Man
Displays a man as infusion in
progress indicator
Enabled / Disabled
Syringe
Displays a syringe as infusion in
progress indicator
Enabled / Disabled
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...