79
8.1
Procedure #1: Battery and Battery Door - Boost Capacitor Discharge
Required Tools and Equipment
1 Torx T-10 screwdriver
1 anti-static bracelet
Maintenance level
Level 3 specialist technician.
The battery must be only replaced by a qualified person.
Dismantling
1.
Unscrew the 2 screws [1] that hold the battery door in place.
2.
Remove the battery door [2].
3.
Power on the pump.
2
1
1