![Fresenius Kabi Agilia SP Скачать руководство пользователя страница 290](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626290.webp)
142
Note: "Ut" is the AC Power voltage prior to applying the test level.
23.4.3
Table 4 - Guidance and Manufacturer's Declaration -
Electromagnetic Immunity
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
- - - - - - - - - - - - - -
400 A/m
If necessary, the power of the magnetic
field should be measured in the intended
installation location to ensure that it is
lower than compliance level.
If the measured field in the location where
Agilia pump is used exceeds the
applicable magnetic field compliance level
above, observe Agilia pump to verify that it
is operating normally. If you notice
abnormal performance, additional
measures may be necessary, such as
reorienting or relocating Agilia pump, or
installing magnetic shielding.
400 A/m
Warning
Agilia pump and its accessories are intended to be used in the
electromagnetic environments specified below.
The customer or the user of Agilia pump should ensure that it is
used in such environments.
Immunity Test
IEC 60601-1-2
- - - - - - - - - - - - - -
Compliance
Level
Obtained by
the Device
Electromagnetic Environment -
Guidance
IEC 60601-2-24
and ANSI/AAMI
ID26
Test Level
Portable and mobile RF communication
equipment should be used no closer to
any part of Agilia pump (including cables),
than the recommended separation
distance calculated from the transmitter
frequency equation.
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
- - - - - - - - - - - - - -
Not applicable
3 Vrms
Recommended separation distance:
D = 0.35
P,
for a frequency of 150 kHz to 80 MHz
Immunity Test
IEC 60601-1-2
- - - - - - - - - - - - - -
Compliance
Level
Obtained by
the Device
Electromagnetic Environment -
Guidance
IEC 60601-2-24
and ANSI/AAMI
ID26
Test Level
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...