![Fresenius Kabi Agilia SP Скачать руководство пользователя страница 63](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626063.webp)
63
6.3
Quality Control Procedure
6.3.1
Running a Quality Control
Print the form provided. See section 6.3.2, page 66.
Run
Test 2: Maintenance
Enter the following maintenance data in the form:
- Running time in hours since last preventive maintenance date
- Last preventive maintenance date
Fix the next preventive maintenance date. Add a maximum of 3 years to the date of last preventive
maintenance.
Fill the Next maintenance date in the form.
1.
Run
Test 1: Identification
. Identification and general appearance check. See section 6.2.2, page 41.
Check housing.
Check AC power cord.
Check the product code
REF
and serial number
SN
on the pump
identification label.
Fill the pump identification data in the form:
- Pump type
- Product code
- Serial number
2.
Check the attachment lock knob and rotating pole clamp (locking system test
).