![Fresenius Kabi Agilia SP Скачать руководство пользователя страница 252](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626252.webp)
104
13.2
Preparing a Syringe
1.
Prepare the fluid to be infused
according to your healthcare facility’s
protocol.
2.
Select a syringe.
3.
Check the syringe and access device
integrity.
4.
Connect the extension set to the
syringe according to local practices.
5.
If necessary, fill the syringe and check
that it is watertight.
6.
Manually prime the extension set
according to your healthcare facility’s
protocol.
7.
Confirm that there is no air in the syringe or in the extension set.
Information
The fluid in the syringe and the syringe must be within normal
operating temperature conditions: +18 °/+30 °C.
It is recommended to prime the set immediately before starting
the infusion.
Do not use in conjunction with positive pressure infusion devices
that could generate back pressure higher than 2000 hPa
(1500 mmHg): doing so will damage the administration set and
the pump.
Some extension sets may have components such as a filter that
require special instructions.
Certain drugs may require specific extension sets.
Connect the infusion line in accordance with procedures in your
healthcare facility using good medical practices. It is
recommended to use a Luer lock system to reduce the risk of
disconnection, leakage, air-in-line, or contamination.
Manually prime the syringe and extension set to remove all air,
before connecting to the pump.
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...