25
4.4.4
Infusion Modes
4.5
Data Set
An infusion can be started according to the following modes:
Infusion Mode
Description
Infusion Rate
Flow Rate
Dose
Simple Rate
Infusion with a programmed rate
Volume/Time
Dose/Time
Infusion of a programmed volume or dose over a
programmed period of time
Volume Limit
Infusion with a limitation on the volume or dose to
be infused
A
data set
is a combination of custom profiles (up to a maximum of 19)
that can be uploaded to Agilia pumps with Agilia Vigilant Drug’Lib
software.
If there is no data set uploaded to the pump, the pump can be used with
the Basic Profile, without the protections of the Agilia Vigilant Drug’Lib
software.
. . .
Device
Configuration
(factory)
Device
Configuration
2
Device
Configuration
3
Device
Configuration
4
Device
Configuration
20
Drug List
B
Drug Library
A
Drug Library
C
Drug Library
S
Data Set
configured with Agilia Vigilant Drug’Lib software
Profile 1
(Basic)
Profile 2
(Custom)
Profile 3
(Custom)
Profile 4
(Custom)
Profile 20
(Custom)
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...