6.2.6
Test 5: Force
Test Description
The syringe force sensor is used to monitor the pressure and detect occlusions in the syringe.
Run this test to validate the accuracy of the force sensor.
This test checks the pressure inside the syringe when an occlusion alarm is triggered.
The force sensor works properly and is correctly calibrated.
This test failed if one of these two conditions is encountered:
The force sensor is not calibrated. Calibrate the force sensor using Agilia Partner
maintenance software.
The force sensor is defective, or not correctly connected to the CPU board. This
problem leads to an error code 22:
- Check the cable connection to CPU board (J9 connector). See figure 2.8,
- Replace the force sensor. See section 8.11, page 110.
Required Tools and Equipment
1 Dynamometer
Prerequisites
The dynamometer’s calibration certificate is valid.
1.
Access the
Tests
menu. See section 6.2.1, page 40.
2.
Press the arrow keys to select
Test 5: Force
.
3.
Press
enter
.
4.
Press the arrow keys to scroll through the information screens
.
The following data is displayed:
Conversion
: Sensor value in mV. The displayed voltage value depends on
the pressure applied on the sensor
Force
: Force value expressed in g
0 sensor
: Sensor offset value expressed in mV. Value saved during last
calibration
Force = 0 kg
: Calibration at 0 kg (0 PSI or 0 bar) in mV. Value saved
during last calibration
Force
=
5 kg
: Calibration at 5 kg (11.6 PSI or 0.8 bar) in mV.Value saved during last calibration
Date
: Date of last calibration displayed in the following format:
Month/Day/Year (mm/dd/yyyy)
Number
: Last calibration number
5.
Press the arrow keys to display the
Conversion/Force
screen.
6.
Push the disengagement lever down and move the plunger driver to the right.
7.
Apply a force to the sensor, then release it.
8.
Check that the
Force
value is between -200 g and +200 g.
CPU BOARD
Force sensor
Main
microprocessor
Force sensor
Integrated
amplifier
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...