Liquiline System CA80SI
Operation
Hauser
87
Outputting the status of a limit switch
Function = Limit switch
Function
Options
Info
Source of data
Selection
Limit switch 1 ... 8
Factory setting
None
Select the limit switch via which the status of the
relay is to be output.
The limit switches are configured in the menu:
Setup
/
Additional functions
/
Limit switches
.
Use the
ALL
and
NONE
soft keys to select
or deselect all the limit switches in one go.
Hold behavior
Selection
• Freeze last value
• Fixed value
• Ignore
Factory setting
Ignore
Outputting the manipulated variable of a controller
To output a controller manipulated variable via a relay, the relay is modulated. The relay is
energized (pulse, t
1
) and is then de-energized (interval, t
0
).
Function = Controller
Function
Options
Info
Source of data
Selection
• None
• Controller 1
• Controller 2
Factory setting
None
‣
Select the controller that should act as the
data source.
Operating mode
Selection
• PWM
• PFM
Factory setting
PWM
PWM=pulse width modulation
PFM=pulse frequency modulation