Operation
Liquiline System CA80SI
78
Hauser
The following information is processed at the binary inputs or output at the binary
outputs:
• Binary inputs:
Signal for process access
: The analyzer can only perform activities that require sample
(measurement, calibration, cleaning) when the signal is active. The time and the
sequence of activities correspond to the settings in the setup. All activities that require
sample are delayed as long as the signal is inactive.
• Binary outputs:
•
Measurement active
signal: Indicates that a measurement is currently running. The
signal is not active in the event of a calibration or cleaning.
•
Sample required
signal: The signal is active for a configurable time before every
activity that requires sample. This makes it possible to activate an external pump or a
dilution module, for example.
1. For the measurement, under
Menu
/
Setup
/
Analyzer
/
Measurement
select
Continuous
as the start condition (analyzer measures continuously, without
interruption, between the measurements).
2. Select
Menu
/
Setup
/
Inputs
/
Binary input
x:y.
3. Configure the binary inputs as follows:
Menu/Setup/Inputs/Binary input x:y
Function
Options
Info
Binary input
Selection
On
Signal type
Selection
Static signal
Signal level
Selection
• High
• Low
Specifies the active signal level:
Low
Input signals between 0 and 5 V DC
High
Input signals between 11 and 30 V DC
1)
x:y = slot no. : input number
4. Link the binary inputs with the analyzer: select
Menu
/
Setup
/
Analyzer
/
Extended
setup
/
Signal for process access
.
5. Select measuring channel
SP1
.
6. Assign a binary input to the selected measuring channel: select
Binary input
x:y.
7. Select
Menu
/
Setup
/
Outputs
/
Binary output
x:y.
8. Configure the binary outputs as follows:
Menu/Setup/Outputs/Binary output x:y
Function
Options
Info
Binary output
Selection
On
Signal type
Selection
Static signal
Function
Selection
Analyzer
Assignments
Function
=
Analyzer
Selection
Measurement active SP1
Here select which binary outputs output the
system status for a measurement currently
running.
1)
x:y = slot no. : input number