Operation
Liquiline System CA80SI
70
Hauser
Menu/Setup/Analyzer/Extended setup/Diagnostics settings/Bottles
Function
Options
Info
Reagent RN
1 to 40 %
Factory setting
5 %
Standard S1
1 to 20 %
Factory setting
2 %
Remaining operating time
Warning limits
Diag. code 726
Reagent RB
1 to 14 d
Factory setting
14 d
Reagent RK
1 to 14 d
Factory setting
14 d
Reagent RN
1 to 14 d
Factory setting
14 d
Standard S1
1 to 14 d
Factory setting
14 d
Alarm limits
Diag. code 727
Reagent RB
1 to 7 d
Factory setting
2 d
Reagent RK
1 to 7 d
Factory setting
2 d
Reagent RN
1 to 7 d
Factory setting
2 d
Standard S1
1 to 7 d
Factory setting
2 d
Absorption curve recording
Menu/Setup/Analyzer/Extended setup/Diagnostics settings/Absorption curve recording
Function
Options
Info
In automatic mode
Selection
• Off
• On
Factory setting
Off
Off
: Recording is only possible in the manual
mode
On
: Recording also takes place in the automatic
mode
Curve
Selection
1 to 7
Factory setting
3
Select the absorption curve to be recorded.
Only one curve can be selected at any one time.
The data are saved in a logbook.