Operation
Liquiline System CA80SI
66
Hauser
5. Use
OK
to confirm or cancel and assign a new file name.
Your setup is saved on the SD card in the "Device" folder.
You cannot upload the exported setup to the device again. You must use the
Save
setup
. This is the only way you can save a setup to an SD card and reload it later on or
upload it to other devices.
Activation code
Activation codes are required for:
• Additional functions, e. g. Fieldbus communication
• Firmware upgrades
If the original device has activation codes, these can be found on the nameplate. The
corresponding device functions are activated at the factory. Codes are only required if
servicing the device.
1. Enter the activation code:
Menu
/
Setup
/
General settings
/
Extended setup
/
Data
management
/
Activation code
.
2. Confirm your entry.
The new hardware or software function is activated and can be configured.
Functions that are enabled by an activation code:
Function
Activation code beginning with
2 current outputs (BASE2-E module only)
081...
Web server
351...
PROFIBUS DP
0B3...
Modbus TCP
0B8...
Modbus RS485
0B5...
EtherNet/IP
0B9...
PROFINET
0B7...
Cooling
0F1...
Feedforward control
220...
Chemoclean Plus
25...
Collecting vessel
20...
Measuring channels
28...
Formula
321...
1)
Via Ethernet socket of the BASE2 module, for versions without an Ethernet fieldbus
2)
3)
Availability depends on measuring parameter
4)
Mathematical function
Changing the password
You can lock the operating keys with a password (access the context menu by pressing the
navigator for longer). The keys can then only be enabled by entering the right password.
You can set the password to lock the keys here:
Menu
/
Setup
/
General settings
/
Extended
setup
/
Data management
/
Change key lock password
.
1. Enter the current password (factory setting 0000).
Enter new password
2. Enter a new password.
Confirm new password