Diagnostics and troubleshooting
Liquiline System CA80SI
136
Hauser
DIAG/System testAnalyzer
Function
Options
Info
Stop
Sample channel
Current channel
Displays the active measuring channel
Flow rate
Displays the current flow of the active measuring
channel
Test channel
Select the measuring channel to be tested
After making the selection, press
Confirm
Confirm
Dispensers
Dispenser selection
Selection
• None
• 1
• 2
• 3
Empty dispenser
Dispenser is emptied
Draw up dispenser
Dispenser draws in reagent from the reagent
bottles
11.13.2 Power supply
DIAG/System test/Power supply
Function
Options
Info
Digital supply 1: 1.2V
Read only
Detailed list of power supply to instrument.
The actual values can vary without a
malfunction having occurred.
Digital supply 2: 3.3V
Analog supply: 12.5V
Sensor supply: 24V
Temperature
11.14 Resetting
DIAG/Reset
Function
Options
Info
Device restart
Selection
• OK
• ESC
Restart and keep all the settings
Factory default
Selection
• OK
• ESC
Restart with factory settings
Settings that have not been saved are lost.